Fuse Replacement
WARNING: ONLY A QUALIFIED SERVICE TECHNICIAN MAY REPLACE SYSTEM FUSES.
1. If the Sapphire
®
Plus source fails to operate after power is applied, disconnect the power cord from AC wall
outlet before checking the fuse(s). Switch the power switch to the OFF (O) position.
2. Check for blown fuses by removing the fuse covers using a flat blade screw driver (medium size) or equivalent.
3. Replace the blown fuse(s) with the same size and rating. Fuses are 6 amp AGC-6 1/4” x 11/4” Fast Acting.
4. Replace fuse cover.
5. Reconnect the power cord.
6. Switch the power switch to the ON ( I ) position.
7. If the unit fails to operate properly again, contact Den-Mat
®
Holdings, LLC for repair at 1-800-445-0345.
Transportation
When transporting the Sapphire Plus, remove the power cord and the light guide from the Sapphire Plus
light source. Hold the handle or carry it with both hands, one hand underneath the unit to insure safety.
To prevent damage to the Sapphire Plus during transportation, it is highly recommended that the unit be wrapped
with bubble wrap and placed in a shipping box or similar packing materials.
Storage
Store the Sapphire Plus and accessories in a clean area and maintain the room temperature as follows:
Storage Environment (Operating): Ambient Temperature 6˚ C to 45˚ C (43˚ F to 113˚ F)
Transportation & Storage Environment (Non-operating): Ambient Temperature -50˚ C to 70˚ C
(-122˚ F to 158˚ F) Relative Humidity 10-95% at 35˚ C (95˚ F)
DO NOT bend the AC cord severely, apply excessive force, pull, twist, or squeeze the power cable.
DO NOT bend the light guide severely, apply excessive force, pull, twist, or squeeze the light guide.
DO NOT subject the Sapphire Plus to excess impact during transportation; this may cause the unit
to malfunction.
AVOID placing the Sapphire Plus in direct sunlight or high temperature.
The Den-Mat Holdings, LLC System Specification
The light source is Class I: The Den-Mat Holdings, LLC light source relies upon the connection to the
protective earth conductor of the installation to prevent shock hazards.
The light source is Type BF: The Den-Mat Holdings, LLC light source’s applied part is floating from earth.
Classification of the light source against ingress of liquids: Ordinary Equipment (IPX0) – This Den-Mat
Holdings, LLC light source provides no protection against the ingress of water.
Classification against Flammable Vapors with Air: This Den-Mat Holdings, LLC light source is not
suitable for use in the presence of a Flammable Anaesthetic Mixture with Air, Oxygen or Nitrous oxide.
System Electromagnetic Information for the Sapphire Light Source Unit (per IEC 60601-1-2)
The Sapphire Plus Light’s RF emissions are very low and are not likely to cause interference in nearby
electrical equipment.
The Sapphire Light (Class B) is suitable for use in all establishments, including those directly connected to
the public low-voltage power supply network that supplies buildings used for domestic purposes.