STM 3W Diode Laser Mode (continued)
The STM 3W Diode Laser unit may be deactivated at any time by
pressing the Emergency “STOP” button on the STM 3W Diode
Laser unit control panel. This action deactivates the unit function
in any mode, at any power setting, and whether or not the Laser
working power is active. To reset the unit for use, depress and
hold down the “STOP” button for 5 seconds; this places the unit
in STAND-BY Mode.
Sanitizing & Sterilizing
Sterilizing The Whitening Crystal
Clean with a mild, non-abrasive, liquid soap and thoroughly dry all surfaces. Cold sterilize with solutions
containing glutaraldehyde, low concentrations of hydrogen peroxide, isopropyl alcohol or other disinfectants
containing mostly inert ingredients.
Sterilizing The Curing Tip
Clean the tip regularly with composite solvent. The removable tip must be sterilized using standard autoclave
procedures for stainless steel instruments (135º C). Use a sterilization bag and leave sealed for future use.
Failure to sterilize tips between patients increases the risk of patient-to-patient infection.
Sterilizing The Single Use/Disposable STM Tip
Cold sterilize with solutions containing glutaraldehyde, hydrogen peroxide, isopropyl alcohol or other
disinfectants containing mostly inert ingredients.
CAUTION: DO NOT AUTOCLAVE THE LIGHT GUIDE!
Sanitizing The Light Guide and Handle
The light guide cannot be autoclaved. It should be sanitized by using a non-abrasive disinfectant. Do not use
glutaraldehyde-type disinfectants.
Cleaning The Sapphire Plus System
NOTE: Unplug the power cord prior to cleaning with any liquid. Allow the system to cool prior to cleaning.
Avoid spilling any liquid on the system. Do not use flammable liquids to clean the main system or light guide.
If the internal electronics are exposed to any liquid, disconnect the power. Do not attempt to start the system.
Contact the factory for instructions at 1-800-445-0345.
Clean the main system by using a mild, non-abrasive disinfectant. Do not saturate the control panel with liquid.
Wipe with a damp towel and dry completely.
Replacement Parts
Additional tips, shields, and curing rings, light guides and lamp modules are available from Den-Mat
®
Holdings,
LLC. See back cover for ordering information.